90M9

90m9 Invasive Anode Current Sensor mA mAsThe 90M9-AG and 90M9 are invasive sensors.

The mAs sensor connects to models Accu-Gold / Rapid-Gold / Accu-Pro / Accu-kV and to the X-ray generator with safety-shrouded banana jacks. The jacks are not polarity sensitive. It is intended to be inserted in series with the X-ray generator return.

The Accu-Pro and Accu-kV require a kV sensor in-beam as a trigger source.

The 90M9-AG and 90M9 measure DC current. When measuring AC current, we recommend using Model 90M9-BR. Click on the “Sensors and Accessories” tab above for details

90M10

90M10 Non-Invasive Anode Current Sensor mA mAsThe 90M10-AG and 90M10 are Non-Invasive Clamp on Sensors.

The mAs sensor connects to models Accu-Gold / Rapid-Gold / Accu-Pro / Accu-kV and couples to the X-ray generator by clamping to the X-ray anode HV cable (clamps up to 23mm diameter).

The Accu-Pro and Accu-kV require a kV sensor in-beam as a trigger source.

To measure Anode current, clamp on the anode cable with the arrow on the clamp pointing towards the X-ray tube. For best results, do not move clamp before and during a measurement or place next to the X-ray tube.